DYNAMIC PASS BOX
Our World class range of Dynamic pass
boxes are designed in conformation to
the international standards and are
used for transferring the material
through controlled environment to
different classified area to avoid cross
contamination & maintain integrity of
product & process.
Description
Description – Dynamic Pass Box (MOC – SS 304, 316, 316L)
HEPA Filter | Imported Minipleat HEPA filter with hot melt technology , which conforms to EU-14 Grade at supply position with an efficiency rating better than 99.999% for 0.3 micron |
---|---|
Pre-filter | Pre-filters are made from Non-woven synthetic with HDPE mesh;conforms to EU-06 Grade, with an efficiency rating better than 95% down to 5 micron |
Fresh Air Filter | Fresh Air filters are made from Non-Woven Synthetic with HDPE mesh; conformsto EU 4 Grade, with an efficiency rating better than 90% down to 10 micron |
Motor blower assembly |
Motor blower designed of statically” and dynamically balanced for supply of sufficient capacity and static pressure to take care of air flow requirement for entire life of HEPA filter |
Pressure Gauge | Magnehelic Gauge (0 – 25 mm: 1 Nos.)to monitor pressure drop across HEPA filter |
Door | SS 304 double walled door construction with flush toughned glass view panel swith SS handles and SS hinges. |
Interlocking | Electro- Magnetic Interlocking type door arrangement to ensure that only one side door can be opened at an given time. |
Door Indication | Indicators on either side for door open condition. |
Door release | Door Release Switch on e either side |
SS Coving | Internal covings for easy cleaning |
ON/OFF Switches | Keypad ON / OFF Switches for motor & Iights |
DOP Port | DOP test port at upstream of HEPA filter integrity test.( WithPAOSolution) |
Vibration Level | Minimum |
Electrical Supply | 230VAC 1- Ø50 HZ / 415V AC 3-Ø 50 HZ |
Noise Level | Less than 67 dB on scale ‘A’ when ambient is greater than 55 dB on scale ‘A’ |
Optional Accessories
1) U.V. Light with Hour Meter
2) Electro-Magnetic / Mechanical Interlocking
3) Floor mount construction with three side wall structure & Door Drop Seal
4) Audio Visual Alarm indicating condition of HEPA Filters
5) Clean down time with operation hold indicator
6) Photohelic gauge / Digital pressure gauge to monitor of pressure drop across filter
7) Flame Proof fitting
Reviews
There are no reviews yet.